Potential Risk in Combined Use of Capsule Retractors and CTR

Show Description +

Altan A. Ozcan, MD, presents a complication associated with capsular tension ring (CTR) implantation after placement of capsule retractors. During attempted removal of one capsule retractor, it was discovered that the CTR had passed through the retractor's distal loop. However, gentle manipulation and rotation of the CTR through the distal loop of the retractor resulted with atraumatic removal of the capsule retractor.

Posted: 1/03/2018

Potential Risk in Combined Use of Capsule Retractors and CTR

Altan A. Ozcan, MD, presents a complication associated with capsular tension ring (CTR) implantation after placement of capsule retractors. During attempted removal of one capsule retractor, it was discovered that the CTR had passed through the retractor's distal loop. However, gentle manipulation and rotation of the CTR through the distal loop of the retractor resulted with atraumatic removal of the capsule retractor.

Posted: 1/03/2018

Please log in to leave a comment.

Comments

Lulu Cheng

6 years ago

Why not just cut the Iris retractor?

fersat muhacir

6 years ago

Thanks for sharing this potential unwanted complication, and congratulations for your perfect surgical manipulations in resolving the problem.